3 RANGE ATTENUATOR.
1000V INPUT CAT III
LOW BATT. AUTO OFF.
SEPARATING DESIGN.CONVENIENT, DURABLE.
MEET CE, IEC 1010, EN 61010, UL 3111.
Band width (-3dB, 50Ω load): DC-25MHz . ( x 20: DC~15MHz) Accuracy : ± 2% at 20-30℃ 70% RH after 20 minutes warm up. Attenuation : x 20, x 50, x 200 ( Into 1MΩ scope ) ; x 40, x 10 , x 400 ( Into 50Ω load ) Maximum operation Voltage ( DC + peak AC ) : ≦ ± 140V at x 20 ≦ ± 350V at x 50 ≦ ± 1300V at x 200 Maximum input Differential Voltage : 1300V ( DC + peak AC ) or 1000V RMS Maximum input Voltage to Ground : 1000V ( DC + peak AC ) or 600V RMS Common Mode Rejection Ratio ( CMRR ): 60Hz : >10,000 : 1 100Hz: >1,000 : 1 1MHz: > 300 : 1 Noise :≦ 2 mV rms ( Into 50Ω load ) Input Impedance : 2MΩ, 2.3 PF between inputs and ground. 4MΩ, 1.2 PF between inputs. Power : external 6V-9V DC.
WARNING
1. Do not use DP-25 above 1000v ( DC + peak AC ) between ground and the input or 1,300V ( DC + peak AC ) between the input lead.
2. Do not operate DP-25 in wet or damp condition.
3. Do not operate DP-25 in an explosive atmosphere.
4. Do not immerse DP-25 in liquids.
5. Do not operated DP-25 without covers.
6. Please change the battery when the "LOW BATT " LED is lighted. At this time DP-25 can operate but not guaranteed the accuracy.
7. DP-25 can not operate if both POWER and LOW BATT LED are not light.
FEATURES
1. The DP-25 FET input differential probe provieds a safe means of measuring circuits with floating potentials up to 1000 V ( DC+ peak AC ) from ground and 1300V ( DC + peak AC ) differential.
2. The DP-25 converts the high voltage differencial input signal to a low voltage ground refereced signal for display on any Oscilloscope.
3. The output BNC of DP-25 is calibrted to drive a high impedance ( 1MΩ ) load.
INSTRUCTION FOR USE
1. Connect the output BNC of DP-25 to the input BNC of the Oscilloscope by the accessory BNC cable.
2. Adjust the vertical offset of the Oscilloscope if necessary.
3. Set the select proper range of the DP-25 and the V / DIV of the Oscilloscope according to the scale conversion chart.
4. Scale conversion chart : The effective V / DIV is the attenuation factor of x 20, x 50, x 200 multiplied by the scale factor of the Oscilloscope. It will be twice when the 50 Ω load was used. For example, with the range set at x 200, and the scope set to 0.5 V / DIV, the effective V / DIV equals 200 x 0.5 or 100 V, with the 50 Ω load was used, it becomes 200V, the power consumption will increase too. NOTE : If the voltage of the input signal exceeeds the linear range of the setting range.The signal output of the DP-25 would not accurately, the wave form display will be cut off.