Monarch Instrument ACT-1B Tachometer

Có: 0 sản phẩm trong giỏ

Monarch Instrument ACT-1B Tachometer
  • Mã SP: MONARCH INSTRUMENT ACT-1B TACHOMETER
  • Giá NY: Liên hệ
  • Giá bán: Liên hệ
  • Tình trạng:
  • Khuyến mãi:
  • Lượt xem: 5117


  • Universal Input
    200 mV to 50 VAC Input
    1/8 DIN
    5 to 99,999 RPM
    115/230 VAC or 12/24 VDC
    Optional Analog Output

  • Tags: Bàn Map
Đơn vị tiền tệ
  • Số lượng:
  • Thành tiền
Specifications: 
  • Range: 5 RPM to 99,999 RPM
  • Resolution: + 1 RPM
  • Power Supply:
    • AC Power: 115 or 230 Vac + 10% 50/60 Hz (Specify when ordering)
    • DC Option: 12 Vdc at 1.75 Watts
  • "Universal Input"
    • AC Signal - Trigger or Wave from 200mV to 50V
    • TTL input 
  • Analog Output: 4 to 20 mA dc with a maximum external excitation voltage of 40 Volts dc or 0 to 5 Vdc. External Power Source Required
    • Linearity: < 0.5%
    • Resolution: 12 bits or 4096 steps (.025%)

Additional Information:

Speed Range:
5 to 99,999 RPM
Accuracy:
±1 RPM or 0.005% of reading
Resolution:
1 RPM
Display:
5 digit, 0.56" (14 mm) high red LED.
Display Update:
Twice per second above 120 RPM. Programmable with PM Remote Software and USB programming cable.
Dimensions:
1/8 DIN by 4.5" (114 mm) deep. Panel cutout: 3.58" x 1.74" (91 x 44mm)
Input Power:
Standard: 100 - 240 Vac ±10%, 50/60 Hz      Optional: 12 or 24 Vdc ±10%, Isolated, 6.0 Watts.
Inputs:
Universal input for optical, proximity, two wire or three wire magnetic, infrared or laser sensors. TTL input or 1.5 Vac to 50 Vac
input. Standard input is 1 pulse per revolution.
Sensor Excitation:
5Vdc or optional 12Vdc or 24Vdc to sensor.
Recommended Sensors:
Optical - Monarch ROS-W, ROS-P-25, ROS-HT-W-25, ROSM-5W
Proximity - Monarch P5-11
Magnetic - Monarch M-190W or MT-190W
Infrared - Monarch IRS-W
Laser - Monarch ROLS-W, SLS-115/230
Inductive - Gas Engine GE-200
Analog Output:
(Optional) 
VOLTAGE (AO): 0 to 5 Vdc out, 5 mA 16 bit resolution. Full scale RPM settings as specified when ordered or programmable using PM Remote Software and USB programming cable. or CURRENT (IO): 4 to 20 mA out, 16 bit resolution 12Vdc compliance voltage. Full scale RPM settings as specified when ordered or programmable using PM Remote Software and USB programming cable.
Pulse Output:
(Optional)
0 to 5 V TTL pulse, non-inverting, one pulse out for each pulse in. Positive or negative out. Programmable using PM
Remote Software and USB programming cable.
Remote Programming:
Unit can be remotely programmed using optional communications interface cable and PM Remote Software.
Environmental:
Indoor use only, Installation category II per IEC 664 Temperature: -10°C to 50°C operating per IEC 61010-1 Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C.
Electrical Safety:
Meets EN61010-1:2001, EC low voltage directive 72/23/EEC and 89/336/EEC
 
Please make a selection from our Remote Tachometer Sensors.
Link Exchange: Duong Ren - Kinh Hien Vi - Panme - Ban Map - Thuoc Cap-