- Hãng sản xuất: Sonel
- Model: MZC-305
- Xuất xứ: Ba Lan
Measurement of short circuit loop impedance (ZL-PE, ZL-N, ZL-L)
Measuring range according to IEC61557: 0,13…1999? for 1,2m test leads; 0,25…1999? for UNI-SCHUKO adapter
Range
Resolution
Accuracy
0,00...19,99Ω
0,01Ω
±(5% m.v. + 3 digits)
20,0...199,9Ω
0,1Ω
±(4% m.v.. + 3 digits)
200...1999Ω
1Ω
• rated operating voltage UnL-N/ UnL-L: 110/190V, 115/200V, 127/220V, 220/380V, 230/400V, 240/415V, 290/500V, 400/690V, • operating voltage range: 100…440V (for ZL-PE and ZL-N) and 100…750V (for ZL-L) • high current measurement (up to 30A), • operating frequency range: 45…65Hz, • control of correctness of PE terminal connection by means of a contact electrode (applicable to ZL-PE), • measurement is also possible when L and N circuits are swapped, • measurement of resistance and reactance components, • short circuit current calculation.
Measurement of short circuit loop impedance ZL-PE RCD (without RCD tripping) Measuring range according to IEC61557: 0,43...1999? for 1,2m test leads; 0,50...1999? for UNI-SCHUKO adapter
±(6% m.v. + 10 digits)
±(6% m.v. + 5 digits)
• operating voltage range: 100 ... 440V, • rated operating voltage Un: 110V, 115V, 127V, 220V, 230V, 240V, 390V, 400V, • control of correctness of PE terminal connection by means of a contact electrode, • measurement is also possible when L and N circuits are swapped, • measurement of resistance and reactance components, • short circuit current calculation.
Voltage measurement
0,0V...249,9V
0,1V
±(2% m.v. + 4 digits)
250V...750V
1V
±(2% m.v. + 2 digits)
• measurement available for all settings of short-circuit loop (ZL-PE, ZL-PE RCD and ZL-N,L-L), • frequency range 45…65Hz, • overvoltage alarm.
Electrical safety: - type of insulation: double, according to IEC 61010-1 - measurement category: CAT IV 600V (CAT III 1000V) - protection class acc. to EN 60529: IP54 Other technical data - power supply: alkaline batteries or NiMH rechargeable batteries, size AA (4 pcs) - number of Z measurements (for alkaline batteries): >3000 (2 measurements per minute) - time to Auto-OFF: 120 seconds - data transmission: USB - memory of measurement results: 990 cells, 3500 entries - display: LCD segment - dimensions: 260x190x60mm - meter weight: ca. 2,2 kg Rated operational conditions: - operating temperature: 0...+45°C - storage temperature: -20...60°C - humidity 20-80% The instrument meets the requirements of EMC standards